Nanophotometer

Implen NanoPhotometer NP80 (IMP-NANO-NP80)

Implen NanoPhotometer NP80

Features

Microvolume & Cuvette

Accurate
Patented Sample Compression TechnologyTM with guaranteed fixed path lengths allows for unmatched accuracy and precision. Independent from surface tension and free from evaporation. Lifetime accuracy guaranteed.
 
Recalibration-free
Sealed optics without path length drift eliminates the need for costly and time consuming recalibrations. Durable, inert surfaces provide easy to clean, scratch resistant surfaces that do not require reconditioning.
 
Easy
NPOS Operating System: intuitive graphical user interfaces (GUI) providing one-step method access, pre-programmed and customizable applications based on the ultimate level of data security.
 
Flexible
Control your NanoPhotometer® via Touchscreen / Smartphone / Tablet / PC (see page 15). Equipped with Wi-Fi, USB A/B, HDMI, and LAN interface connections. Define and store your own methods.
 
Mobile
Spectroscopy can now be taken anywhere. The optional integrated battery pack provides up to 8 hours of standalone battery powered operation.
 
Powerful
Built-in computer with a high performance Q7 quad core 1 GHz processor and 8 GB of onboard memory provides rapid analysis and easy data storage.
 
Fast
Power on and instantly measure without lamp warm-up time. Full scan capability from 200 - 900 nm provides rapid and complete sample analysis in only 3.5 - 6 sec.

Delivery time: 4-6 Weeks
Delivery times may vary
Request Info
Request Quote
Send To a Friend
Print

NanoVolume Performance

Detection Range dsDNA 1 ng/µl to 16,500 ng/µl

(N50: 5 ng/µl to 7,500 ng/µl)

Detection Range BSA 0.03 mg/ml to 478 mg/ml

(N50: 0.15 mg/ml to 217 mg/ml)

Minimum Sample Size 0.3 µl

Photometric Range

(10 mm equivalent)

0.02 - 330 A

(N50: 0.1 - 150 A)

Path Length 0.67 and 0.07 mm

Dilution Factor 15 and 140

Vortexer 2,800 rpm; tube size up to 2.0 ml

Cuvette Performance

Detection Range dsDNA 0.1 ng / µl to 130 ng / µl

Detection Range BSA 0.003 mg / ml to 3.7 mg / ml

Photometric Range 0 - 2.6 A

Center Height (Z-Height) 8.5 mm

Cell Types Outside dimension 12.5 x 12.5 mm

Heating 37°C ± 0.5°C

Optical Specifications

Wavelength Scan Range 200 – 900 nm (N50: 200 - 650 nm)

Measure Time For Full

Scan Range

3.5 - 6.0 seconds

Wavelength

Reproducibility

± 0.2 nm (N50: ± 1nm)

Wavelength Accuracy ± 0.75 nm (N50: 1.5 nm)

Bandwidth better than 1.8 nm (N50: 5 nm)

Stray Light < 0.5 % at 240 nm using NaI (N50: < 2%) and

< 1% at 280 nm using Acetone (N50: < 2%)

Absorbance

Reproducibility

< 0.002 A (0.67 mm path) @ 260 nm

(N50: 0.004 A (< 0.67 mm path) @ 260 nm)

Absorbance Accuracy < 1.75 % @ 0.7 A (0.67 mm path) @ 260

nm of the reading

Zero Stability ±0.003 A/hour after 20min warm up @ 260 nm

Noise 0.002 A rms at 0 A @ 260 nm

0.002 A (pk to pk) at 0 A @ 260 nm

Optical Arrangement 1 x 3648 CCD Array

(N50: 1 x 1024 CCD Array)

Lamp Xenon flash lamp

Lifetime 109 flashes, up to 10 years

Processing Power & Compatibility

Operating System Linux based NPOS

Onboard Processor Quad Core 1 GHz

Internal Storage 8 GB

Control Options Onboard with built-in Touchscreen,

Computer, Smartphone and Tablet

Software Compatibility Windows 7 (32 & 64 bit), Windows 8 (32 &

64 bit), OS X, iOS & Android OS

Min. Requirement

Smartphone/Tablet

4 inch screen; Apple: iPad 2, iPhone5 &

iOS 6; Android OS version 2.2, Quadcore 1.2

GHz with 1 GB RAM

General Specifications

Main Body Size 200 mm x 200 mm x 120 mm

Weight 3.8 - 5.2 kg depending on configuration

Operating Voltage 90-250 V, 50/60 Hz, 60 W (90 W with

battery pack), 18/19 VDC

Display 1024 x 600 pixels; Touchscreen glove

compatible

Built-in Battery Pack Optional rechargeable lithium ion battery;

95Wh, 6,6Ah; Operation time: up to 8 h;

min. charging cycles: 800

Certification CE, IEC 61010-1:2012 and EN 61326-1:2013

Battery Certification IEC 62133 and UN38.3 transport test

In & Output Ports 2x USB A, USB B, HDMI, Ethernet, WLAN

Additional Data Input Mouse & keyboard options

Security Slot for Kensington lock